Specifications
Chapter 6 45
Making Distortion Measurements
Measuring Harmonics and Harmonic Distortion with a One-Button
Measurement
Making Distortion Measurements
Measuring Harmonics and Harmonic 
Distortion with a One-Button Measurement
This procedure measures the harmonics of the 10 MHz reference output 
signal. The harmonics and total harmonic distortion are measured 
using the one-button automated harmonic measurement.
Step 1. Preset the analyzer:
Press 
Preset, Factory Preset (if present).
Step 2. Connect the ESA 10 MHz reference output from the rear of the 
analyzer to the INPUT. For PSA turn the internal 10 MHz reference 
signal on:
(PSA) Press 
System, Reference, 10MHz Out (On).
Step 3. Set the analyzer reference level, center frequency and RBW:
Press 
AMPLITUDE Y Scale, Ref Level, 10, dBm.
Press 
FREQUENCY Channel, Center Freq, 10, MHz.
Press 
BW/Avg, Res BW, 300, kHz.
Step 4. Run the Power Suite harmonic distortion measurement:
Press 
Measure, More, Harmonic Distortion.
Step 5. Set the number of harmonic distortion measurement averages to 3:
Press 
Meas Setup, Avg Number (On), 3, Enter
Step 6.
Set the average mode to exponential to continuously update the result 
after each subsequent sweep:
Press 
Meas Setup, Avg Mode (Exp).
Repeat average mode clears the averaged result after the specified 
number of averages is complete.
Step 7. Optimize the analyzer’s dynamic range settings:
Press 
Meas Setup, Optimize Ref Level.
Step 8. Display the total harmonic distortion: 
(ESA) Press 
View/Trace, Harmonics & THD.
(ESA) Press 
Trace/View, Harmonics & THD.