Specifications
Chapter 9 77
Measuring Digital Communications Signals
Making Multi-Carrier Power (MCP) Measurements
Measuring Digital Communications 
Signals
Making Multi-Carrier Power (MCP) 
Measurements
The multi-carrier power measurement measures the total power of up 
to 12 carriers and their adjacent channels for up to three pairs of offset 
frequencies. The offset frequency properties can be modified, including 
the offset frequency, the integration bandwidth and upper/lower limits. 
The carrier parameters can also be modified including carrier width, 
carrier integration bandwidth and whether the carrier is on or off. MCP 
can be setup with no radio standard selected or with radio standard 
settings for IS-95 and W-CDMA. Results for carriers without power 
present are displayed relative to the reference carrier. Results for 
adjacent channels are displayed in absolute power (dBm). 
The following example shows how to make an MCP measurement on a 
W-CDMA base station broadcasting with 8 carriers on and with two 
carriers off. The transmitting carriers are spaced at 5 MHz intervals at 
the following frequencies: 977.5 MHz, 982.5 MHz, 987.5 MHz, 
992.5 MHz, 1.0075 GHz, 1.0125 GHz, 1.0175 GHz and 1.0225 GHz. 
NOTE Recommended equipment: The Agilent ESG signal generators can 
be used to transmit W-CDMA multi-carrier signals of up to 4 carriers. 
The multi-carrier signal parameters can be modified including channel 
setup, frequency offset and carrier power. An alternative way to setup 
multiple carriers is to use multiple ESGs, each transmitting one 
W-CDMA signal. 
Step 1. Connect a W-CDMA signal with multiple carriers broadcasting at the 
frequencies stated above.
Step 2. Set the center frequency of the analyzer to the midpoint of all the 
carriers:
Press 
Preset, Factory Preset (if present).
Press 
FREQUENCY Channel, Center Freq, 1, GHz.
Step 3. Set the analyzer radio mode to 3GPP W-CDMA as a base station device:
Press 
Mode Setup, Radio Std, 3GPP W-CDMA.
Press 
Mode Setup, Radio Std Setup, Device (BTS).
Step 4. Select the multi-carrier power one-button measurement from the 
measure menu and then optimize the reference level and attenuation 
settings suitable for the MCP measurement:
Press 
MEASURE, Multi Carrier Power.
Press 
Meas Setup, Optimize Ref Level.
Step 5. Set the carrier number to 10 (in our multi-carrier setup we have 8 
carriers without 2 middle carriers):










