Specifications
78 Chapter 9 
Measuring Digital Communications Signals
Making Multi-Carrier Power (MCP) Measurements
Measuring Digital Communications 
Signals
Press Meas Setup, Carrier Setup, Carriers, 10, Enter.
Step 6. Configure carrier 5 to have no power present:
Press 
Meas Setup, Carrier Setup, Configure Carriers, Carrier, 5, Enter, 
Carrier Pwr Present (No).
Step 7. Repeat step 6, configuring carrier 6 to have no power present.
Step 8. Display the results in full screen view (see Figure 9-9):
Press 
Display, Full Screen.
Figure 9-9 MCP Measurement on a Base Station W-CDMA Signal
In this example, the intermodulation falls outside the transmit 
channels which are marked by the colored vertical lines. The white set 
indicates the reference carrier. The red sets contain the carriers with 
power present and the blue lines mark the carriers without power 
present. Limits for the upper and lower offsets can also be set as shown 
in the example: “Making Adjacent Channel Power (ACP) 
Measurements” on page 74.
NOTE Press the Return key to exit the full screen display without changing 
any parameter values.
Step 9. View the results table of carriers 7-10:
Press 
Meas Setup, Carrier Result, 7, Enter.
Step 10. View the results in a combined spectrum and bar graph (see 
Figure 9-10):
(ESA) Press 
View/Trace, Combined.
(PSA) Press 
Trace/View, Combined. 