Agilent 16048A Test Leads Operation and Service Manual Fourth Edition Agilent Technologies Japan, Ltd.
Notices The information contained in this document is subject to change without notice. This document contains proprietary information that is protected by copyright.All rights are reserved. No part of this document may be photocopied, reproduced, or translated to another language without the prior written consent of the Agilent Technologies. Agilent Technologies Japan, Ltd.
• Ground The Instrument To avoid electric shock hazard, the instrument chassis and cabinet must be connected to a safety earth ground by the supplied power cable with earth blade. • DO NOT Operate In An Explosive Atmosphere Do not operate the instrument in the presence of flammable gasses or fumes. Operation of any electrical instrument in such an environment constitutes a definite safety hazard. • Keep Away From Live Circuits Operating personnel must not remove instrument covers.
Warranty This Agilent Technologies instrument product is warranted against defects in material and workmanship for a period corresponding to the individual warranty periods of its component products. Instruments are warranted for a period of one year. Fixtures and adapters are warranted for a period of 90 days. During the warranty period, Agilent Technologies Company will, at its option, either repair or replace products that prove to be defective.
Assistance Product maintenance agreements and other customer assistance agreements are available for Agilent Technologies products. For any assistance, contact your nearest Agilent Technologies Sales and Service Office. Addresses are provided at the back of this manual. Safety Symbol General definitions of safety symbols used on the instrument or in manuals are listed below.
Contents 1. Operation Product Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Typical Characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Compensation for Fixture Residual Impedance Error . . . . . . . . . . . . . . . . . . . . . 10 11 12 13 2. Service Maintenance . . . . . . .
Contents 8
1 Operation This operating note provides complete information on the 16048A Test Leads. The 16048A is shown pictorially in Figure 1-1, its physical dimensions are given in Table 1-1, and typical characteristics related to offset error are given in Table 1-2. To order additional copies of this operating note, use the part number listed on the rear cover.
Operation Product Description Product Description The 16048A consists of a direct attachment, 4-terminal pair type fixture which is equipped with four BNC (m) terminated-coaxial test leads. These test leads are used to attach user-fabricated test fixtures. DC bias levels of up to ±40V can be applied to the 16048A. Cable length is 1 meter. The 16048A is shown in Figure 1-1.
Operation Specifications Specifications Table 1-1 Specifications of the 16048A Function: For use with 4 terminal-pair LCR Meters and Impedance Analyzers. Connector Type: BNC male Cable Length: Approximately 1m Maximum Voltage: ± 40V peak max. (AC+DC) Weight: 315 g Safety Standards: EN61010-1:1993 +A2:1995 IEC61010-1:1990 +A1:1992 +A2:1995 CSA C22.2 No.1010.
Operation Typical Characteristics Typical Characteristics Table 1-2 Typical Characteristics Incremental error at freq. ³1 MHz Parameter reading error (%) f 2 ± 5 ´ æ ------ö è 10ø NOTE Offset value for D f ± 0.02 ´ æ ------ö è 10ø 2 f is the measurement frequency in megahertz. The incremental errors calculated from the equation in the table for measurements at frequencies above 1 MHz are additive.
Operation Compensation for Fixture Residual Impedance Error Compensation for Fixture Residual Impedance Error The 16048A has inherent stray capacitance, residual inductance, and residual resistance that affect the accuracy of measured values. To compensate for, or negate, these residuals to minimize measurement error, the instrument’s Open/Short compensation procedure should be performed. The procedure is given in the instrument’s operating manual.
Operation Compensation for Fixture Residual Impedance Error 14 Chapter 1
2 Service 15
Service Maintenance Maintenance An exploded view of the 16048A (for parts identification) is shown in Figure 2-1. Do not disassemble any further than shown. Maintenance consists principally of cleaning contacts and replacing worn or damaged parts. Take special care when cleaning contacts. To order parts, use the Agilent Technologies part numbers listed in Table 2-1.
Service Maintenance Figure 2-1 Parts Identification Chapter 2 17
Service Maintenance Table 2-1 Parts Identification Reference 18 Agilent Part No. Qty.
Service Maintenance Reference Agilent Part No. Qty. Description 31 1250-0118 4 CONNECTOR-BNC FEMALE 32 2360-0115 2 SCREW 33 16032-10021 1 PLATE 34 2190-0016 4 WASHER 35 2950-0001 4 NUT 36 16032-10022 1 PLATE 37 16032-60001 1 BNC BRACKET Note 1 thru 30 31thru36 a. Agilent internal-only part number. *: Not separately replaceable.
Service Maintenance 20 Chapter 2
REGIONAL SALES AND SUPPORT OFFICES For more information about Agilent Technologies test and measurement products, applications, services, and for a current sales office listing, visit our web site: http://www.agilent.com/find/tmdir. You can also contact one of the following centers and ask for a test and measurement sales representative. 11/29/99 United States: Agilent Technologies Test and Measurement Call Center P.O.