Technical data
Agilent B1500A Configuration and Connection Guide, Edition 5 3- 25
Connection Guide for Wafer Prober
WGFMU and RSU
These instructions apply when DC probes are used on Prober.
• Connect a SSMC short open cable between the Gate DC probe and the well DC probe, 
and set the DC probes to the appropriate place. Then, the black sleeve plug must be the 
Gate side. This electrically connects the Well probe needle, Well probe shield, and Gate 
probe shield together.
• Connect a SSMC short-open cable between the Drain DC probe and the Source DC 
probe, and set the DC probes to the appropriate place. Then, the black sleeve plug must 
be the Drain side. This electrically connects the Source probe needle, Source probe 
shield, and Drain probe shield together.
• Connect the last SSMC short-open cable between the Well DC probe and the Source 
DC probe, and set the DC probes to the appropriate place. Then, the black sleeve plug 
must be the Source side. This electrically connects the Well probe needle, Well probe 
shield, and Source probe shield together.
• Connect a SMA-SSMC cable between a RSU (ex: RSU1) and the Drain DC probe. 
And set the Drain DC probe to the appropriate place.
• Connect the other SMA-SSMC cable between the other RSU (ex: RSU2) and the Gate 
DC probe. And set the Gate DC probe to the appropriate place.
Figure 3-27 SSMC Short-Open Cable
50 mm or 75 mm
Black
SSMC(plug)
SSMC(plug)
Yellow
Signal line and shield are shorted.
no signal pin