Specifications
3-2 Agilent 4155B/4156B Quick Start Guide, Edition 2
Analyzing the Results
In the previous chapter, you measured the drain current (Id) while performing a 
synchronous sweep of the gate voltage (Vg) and drain voltage (Vd). And the 
measurement results were drawn graphically on the screen.
In this chapter, you analyze the measurement results on the graph and search 
threshold voltage (Vth) of the DUT.
The basic algorithm to search for the threshold voltage is:
1. Assign gate voltage (Vg) to X-axis, ÖId
 to Y1-axis, and ¶ÖId / ¶Vg to Y2-axis.
2. Search for the maximum value of ¶ÖId
 / ¶Vg curve, which is also the point 
where the gradient of ÖId
 curve is maximum.
3. Draw a tangent line to the point where the gradient of ÖId
 curve is maximum.
4. Read the X-coordinate value where the tangent line crosses the X-axis. This 
value is threshold value (Vth).