Technical data
Module 7
Measurement Functions
7-36
Series Resistor
(1 M ohm)
SMU 
Force
Sense
Guard
Sense line is internally disconnected 
during the series resistor is connected
GNDU
Use short cable to minimize 
the guard capacitance 
SMU Series Resistor
The series resistor (approx. 1 Mohm) is mounted on each SMU. The series resistor may be used for 
the device protection, negative resistance measurement, and so on. It depends on the characteristics 
of test device and measurement environment. 
Note that the series resistor may not effective for device protection. For example, the breakdown 
measurement may damage the device even if the series resistor is used.
If you use a series resistor, the voltage you set is applied to the near side of the series resistor. Thus, 
the voltage will be divided by the series resistor and the device under test. Then, the measurement 
data is not compensated.
Restrictions:
The series resistor cannot be used for the measurements that use the HRSMU connected to the ASU 
or the measurements that use 1 A range of the HPSMU.