User`s guide
3-32 Agilent 4155C/4156C User’s Guide Vol.2, Edition 6
Sampling Measurements
Defining Measurement Conditions
To Define Stop Conditions 
The measurement stop condition defines the condition to stop the sampling 
measurement. The stop condition is one of the sampling completion conditions. 
1. Press 
Meas key in the PAGE CONTROL key group.
2. Select SAMPLNG SETUP primary softkey.
3. In the ENABLE/DISABLE field of the STOP CONDITION area, select 
ENABLE secondary softkey.
4. In NAME field of STOP CONDITION area, select the secondary softkey for the 
desired variable name or user function name to be used for EVENT comparison.
5. In THRESHOLD field of STOP CONDITION area, enter the threshold value for 
the name selected in the previous step.
6. In EVENT field of STOP CONDITION area, select:
• Val > Th secondary softkey to stop the sampling when the sampled value is 
greater than the threshold value.
• Val < Th secondary softkey to stop the sampling when the sampled value is 
less than the threshold value.
• |Val| > |Th| secondary softkey to stop the sampling when the absolute 
sampled value is greater than the absolute threshold value.
• |Val| < |Th| secondary softkey to stop the sampling when the absolute 
sampled value is less than the absolute threshold value.
7. In EVENT NO. field, enter a value which specifies the sampling to stop when 
EVENT occurs EVENT NO. times.
If you select DISABLE in the ENABLE/DISABLE field, the sampling 
measurement continues until:
•
Stop key in the MEASUREMENT key group is pressed.
• Specified total sample time has elapsed.
• The 4155C/4156C receives GPIB command to stop sampling.
• An emergency condition occurs on the 4155C/4156C.
• Interlock terminal opens due to high voltage. 