User`s guide
4-16 Agilent 4155C/4156C User’s Guide Vol.2, Edition 6
Quasi-static C-V Measurements
Defining Measurement Conditions
To Display List Results 
1. Press the Display key in the PAGE CONTROL key group.
2. Select the DISPLAY SETUP primary softkey.
3. In the DISPLAY MODE field, select the LIST secondary softkey.
4. In the LIST area, select the secondary softkey of the variables for which you 
want to list the measurement results.
When the pointer is in the NAME row, the permissible variable names appear in the 
secondary softkey area. To set a variable name, select the desired secondary softkey. 
The permissible names are names that you already set up on the 
CHANNEL DEFINITION, USER FUNCTION, USER VARIABLE, and 
QSCV MEASURE SETUP screens.
To display variables on the data variable area
In the DATA VARIABLES fields, select the secondary softkey for the desired 
variable. Up to two variables can be displayed on the data variable area, which is 
located below the list area.
To change resolution of the data to be displayed 
In the DATA DISPLAY RESOLUTION field, select NORMAL for the resolution 
shown in the specifications, or EXTEND for the ADC full scale resolution. For 
example, the data display resolution is 1 fA in NORMAL mode, and 10 aA in 
EXTEND mode for the 10 pA measurement range.