User`s guide
5-4 Agilent 4155C/4156C User’s Guide Vol.2, Edition 6
Stress Force
Stress Function
Stress Output Channels 
Stress output channel is defined as the unit used to force stress.
Available Units
The 4155C/4156C can force dc voltage stress, dc current stress, and ac voltage 
stress (by PGUs in Agilent 41501A/B), but cannot force ac current stress.
Table 5-1 shows available units and allowable modes for stress sources.
Table 5-1 Available Units and Modes for Stress Force 
Also, SMUs can be set to COMMON mode.
Setting the Stress Channels
You can set up units for the stress force state on the STRESS: CHANNEL 
DEFINITION screen independently from the measurement and standby states that 
you set on the CHANNELS: CHANNEL DEFINITION screen.
For example, you can use the same SMU as a measurement channel in the 
measurement state, and as a stress force source in the stress force state.
To set a unit to the stress channel, select SYNC in the FCTN field for the unit. If you 
select NSYNC, the 4155C/4156C does not monitor the bias output time for the unit. 
The unit is called as non-stress channel or bias channel in this manual.
To use the stress force function, at least one unit must be set to the stress channel 
which the FCTN field is set to SYNC. You can select up to four stress source 
channels among SMUs, VSUs, and PGUs.
If a unit is set to STBY ON on the CHANNELS: CHANNEL DEFINITION screen, 
the unit cannot be set to the stress channel.
If you use two PGUs as ac pulse source, both PGUs must be the stress channel or the 
non-stress channel. This means that if PGU1 is a stress channel, PGU2 must be a 
stress channel, not non-stress channel.
Units  DC Voltage Stress  DC Current Stress 
AC Voltage Stress 
(Pulsed Source)
SMU yes yes
VSU yes
PGU yes yes










