User`s guide
6-22 Agilent 4155C/4156C User’s Guide Vol.2, Edition 6
Analyzing Measurement Results
Manual Analysis
To Draw Line through Specified Point with Specified 
Gradient 
1. Select LINE primary softkey.
2. Set LINE SELECT softkey to 1 or 2. Selecting this softkey toggles the setting.
3. Set LINE secondary softkey to ON. A line and two cursors are displayed. 
Selecting LINE secondary softkey toggles as follows:
OFF → ON → OFF
4. Select GRAD MODE secondary softkey if it is not highlighted. Softkey 
becomes highlighted. One cursor disappears (if there are two cursors in the 
plotting area before this step). Selecting GRAD MODE secondary softkey 
toggles between highlighted and not highlighted.
5. Move the cursor to desired location by using arrow keys of the 
MARKER/CURSOR key group.
6. Select GRAD VALUE secondary softkey, then enter gradient value. The line 
goes through the cursor with specified gradient.
When lines are displayed and when ON is set in the LINE PARAMETER field on the 
DISPLAY: DISPLAY SETUP screen, the X and Y intercepts and gradients of 
selected line are also displayed in the plotting area.
To turn off the line intercept and gradient display
Select DISPLAY SETUP primary softkey, then set LINE PRMTRS secondary 
softkey to OFF.
To turn off the data variable display area
Use the following procedure:
1. Select DISPLAY SETUP primary softkey.
2. Set DATA VAR secondary softkey to OFF.
To move the selected cursor to the selected marker position
Select CURSOR TO MARKER secondary softkey.