User`s guide
Agilent 4155C/4156C User’s Guide Vol.2, Edition 6 1-5
Sweep Measurements
Measurement Functions
Figure 1-1 Basic Sweep Measurement 
To set up basic sweep measurement, select VAR1 function for desired SMU or VSU 
on CHANNELS: CHANNEL DEFINITION screen.
Parameters
Also, specify the following parameters for VAR1 on MEASURE: SWEEP SETUP 
screen.
Parameter  Description
sweep mode  Single or double sweep.
linear/log  Linear or logarithmic sweep. For logarithmic sweep, select the 
number steps in one decade as follows:
LOG10  10 steps in one decade.
LOG25  25 steps in one decade.
LOG50  50 steps in one decade.