User`s guide
1-6 Agilent 4155C/4156C User’s Guide Vol.2, Edition 6
Sweep Measurements
Measurement Functions
start  Start value of sweep. For logarithmic sweep, start must not be 
zero. Allowable range of start depends on output range of 
sweep source. For output range of each measurement channel, 
refer to Chapter 7.
stop  Stop value of single sweep or turning back value of double 
sweep. For logarithmic sweep, stop must have same polarity as 
start, and must not be zero. Allowable range of stop depends on 
output range of sweep source. For output range of each 
measurement channel, refer to Chapter 7.
step • For linear sweep, step is step increment of sweep. Number 
of sweep steps is calculated from start, stop, and step. 
Calculated number of steps must be in range: 2 to 1001.
• For logarithmic sweep, step is invalid. Number of sweep 
steps is calculated from start, stop, and number of steps in 
one decade, which is specified by log parameter. Calculated 
number of steps must be in range: 2 to 1001.
compliance  Compliance value of sweep source. This parameter applies to 
SMU only. Allowable range of compliance depends on the 
compliance range of sweep source. For the compliance range of 
each measurement channel, refer to Chapter 7.
power compliance  (Optional) Power compliance value of sweep source. This 
parameter applies to SMU only. Allowable range depends on 
power compliance range of sweep source. For details, refer to 
Chapter 7.
hold time  Time required for DUT to settle after forcing start value. 
Allowable range is 0 to 655.35s. Resolution: 10 ms.
delay time  Time required for DUT to settle after stepping the output. 
Allowable range: 0 to 65.535 s. Resolution: 100 μs
Refer to Setup Screen Reference for setting up these parameters.