User`s guide
7-10 Agilent 4155C/4156C User’s Guide Vol.2, Edition 6
Measurement Units and Functions
Measurement Units
Table 7-6 MPSMU Measurement Voltage Values and Resolutions 
Table 7-7 MPSMU Output Current Ranges and Resolutions 
Range 
Measurement 
Value 
a
a. This column is applied to the auto ranging or the limited auto ranging. For fixed ranging, maximum mea-
surement value is Range column value.
Measurement Resolutions 
b
b. Measurement resolution depends on the integration time setting. For Knob sweep measurement, see the 
column of Integration Time 80 μs to 560 μs.
Integration Time 
High Speed 
Sampling 
Measurement 
c
c. This column is applied to the sampling measurement that initial interval is set to 480 μs or shorter. 
1PLC or 
Longer 
640 μs to 1.92 
ms 
80 μs to 560 μs 
2 V 0 ≤ |V| ≤ 2.2 V 2 μV20 μV200 μV2 mV
20 V 0 ≤ |V| ≤ 22 V 20 μV200 μV 2 mV 20 mV
40 V 0 ≤ |V| ≤ 44 V 40 μV400 μV 4 mV 40 mV
100 V 0 ≤ |V| ≤ 100 V 100 μV 1 mV 10 mV 100 mV
Range Output Value 
Output 
Resolution 
Voltage 
Compliance 
Range 
1 nA 0 ≤ |I| ≤ 1 nA 100 fA
±100 V
10 nA 0 ≤ |I| ≤ 10 nA 1 pA
±100 V
100 nA 0 ≤ |I| ≤ 100 nA 10 pA
±100 V
1 μA0 ≤ |I| ≤ 1 μA100 pA
±100 V
10 μA0 ≤ |I| ≤ 10 μA1 nA
±100 V
100 μA0 ≤ |I| ≤ 100 μA 10 nA
±100 V
1 mA 0 ≤ |I| ≤ 1 mA 100 nA
±100 V
10 mA 0 ≤ |I| ≤ 10 mA 1 μA
±100 V
100 mA 0 ≤ |I| ≤ 20 mA 10 μA
±100 V
20 mA < |I| ≤ 50 mA 10 μA
±40 V
50 mA < |I| ≤ 100 mA 10 μA
±20 V