Agilent 4155C/4156C User’s Guide Vol.2, Edition 6 7-29
Measurement Units and Functions
Measurement Ranging Mode
Measurement Ranging Mode 
Before executing measurements, you select a ranging mode from the following four 
modes. You can set the ranging mode for each measurement unit.
• “Auto Ranging”
• “Limited Auto Ranging”
• “Compliance Range”
• “Fixed Range”
The following table lists the allowable measurement ranging modes for each 
measurement mode.
Table 7-21 Allowable Measurement Ranging Modes 
If you choose sweep measurement or sampling measurement (initial interval ≥ 2 ms) 
and you do not set a ranging mode, auto ranging is set for voltage measurement 
mode channel and limited auto (1nA) ranging is set for current measurement mode 
channel automatically.
For sampling measurement (initial interval < 2 ms), an error occurs if fixed range is 
not set.
Measurement Mode  Ranging Mode
Sweep Measurement Auto, Limited Auto, Fixed
Sampling Measurement 
(Initial Interval 
a
 ≥ 2 ms)
a. You specify initial interval on the MEASURE: SAMPLING SETUP 
screen.
Auto, Limited Auto, Fixed
Sampling Measurement 
(Initial Interval 
a
 < 2 ms)
Fixed
Knob Sweep Compliance