User`s guide
1-10 Agilent 4155C/4156C User’s Guide Vol.2, Edition 6
Sweep Measurements
Measurement Functions
Parameters
The parameters for primary sweep source (VAR1) are same as the parameters for 
sweep source of basic sweep measurement. For synchronous sweep source (VAR1'), 
specify the following parameters on MEASURE: SWEEP SETUP screen.
Parameter Description
offset  Offset between outputs of primary and synchronous sweep 
sources.
ratio  Ratio between outputs of primary and synchronous sweep 
sources.
compliance  Compliance value of synchronous sweep source. This 
parameter applies to SMU only. Allowable range of compliance 
depends on the compliance range of synchronous sweep source. 
For the compliance range of each measurement channel, refer to 
Chapter 7.
power compliance  (Optional) Power compliance value of synchronous sweep 
source. This parameter applies to SMU only. Allowable range 
of power compliance depends on the power compliance range 
of synchronous sweep source. For details, refer to Chapter 7.
The relationship between the output of primary and synchronous sweep sources is 
determined by the following equation:
synchronous output = primary output × ratio + offset
The synchronous output determined by above equation must not exceed the output 
range of synchronous sweep source.