User`s guide
7-36 Agilent 4155C/4156C User’s Guide Vol.2, Edition 6
Measurement Units and Functions
Measurement Time
Table 7-24 Integration Time in Short Mode (Integ time=0.96 to 1.92 ms)
Measurement 
Range 
Integration 
Time
Unit and Voltage Range 
a
a. Voltage range is the output range used to force voltage.
10 pA 10 PLC HRSMU
100 pA  2 PLC HRSMU
1 nA  1 PLC HRSMU
Setting value MPSMU/HPSMU using 40 V range or less
Setting value × 2 MPSMU/HPSMU using 100 V range
Setting value × 4 HPSMU using 200 V range
10 nA Setting value × 4HRSMU
Setting value  MPSMU / HPSMU
100 nA Setting value × 2 HRSMU using 2 or 20 V range
Setting value × 4 HRSMU using 40 or 100 V range
Setting value  MPSMU/HPSMU using 2 or 20 V range
Setting value × 2 MPSMU/HPSMU using 40 V range
Setting value × 4 MPSMU/HPSMU using 100 V range
Setting value × 8 HPSMU using 200 V range
1 μA Setting value  HRSMU / MPSMU
Setting value  HPSMU using 100 V range or less
Setting value × 2 HPSMU using 200 V range
10 μA Setting value  HRSMU using 2 V range
Setting value × 2 HRSMU using 20 V range or more
Setting value  MPSMU/HPSMU using 2 or 20 V range
Setting value × 2 MPSMU/HPSMU using 40 V range
Setting value × 3 MPSMU/HPSMU using 100 V range
Setting value × 6 HPSMU using 200 V range
100 μA to 
100 mA
Setting value  HRSMU / MPSMU / HPSMU
1 A Setting value  HPSMU










