User`s guide
Agilent 4155C/4156C User’s Guide Vol.2, Edition 6 1-11
Sweep Measurements
Measurement Functions
Pulse Sweep Measurement 
For a sweep measurement, a sweep or constant source SMU can be a pulse source. 
But only one SMU can be a pulse source. Figure 1-4 shows the relationship between 
pulse source and other sources.
Figure 1-4 Pulse Source and Other Sources 
For the pulse sweep measurement, the delay time of the primary sweep source is 
ignored, and each step of the primary sweep source is synchronized with output of 
the SMU pulse source. Measurements are made during the pulse output.
The pulse output of PGU is not synchronized with any other source.