User`s guide
10-6 Agilent 4155C/4156C User’s Guide Vol.2, Edition 6
Connecting Measurement Devices
Using Connector Plate
Example 
The following example connection can be used to reduce the leakage current. 
Extend the outer conductor as close as possible to the probing needle. This also 
reduces the induced noise.
Guarding 
Guarding reduces the leakage current between the measurement points and 
instrument. This is important when you measure low current.
The following figure shows the theory of guarding. The buffer amplifier (×1) keeps 
the potential of the guard conductor at the same potential as the force conductor, so 
current does not flow between the force and guard conductors. Therefore, the 
current measured by SMU is same as current at measurement point because no 
current is leaked.