User`s guide
Agilent 4155C/4156C User’s Guide Vol.2, Edition 6 3-7
Sampling Measurements
Measurement Functions
Sampling Measurement Data
Measurement parameters of sampling measurement are set to the NAME column of 
the DISPLAY: DISPLAY SETUP screen. Available parameters and example 
parameters for the NAME field are listed in the table below:
Measurement start time can be expressed by the following fomula. This formula is 
available for the measurement points before starting the discarding operation for the 
linear sampling or thinned-out sampling. For logarithmic sampling, this is available 
for the measurement points in the first decade.
@TIME = Thold + Tinterval × [(@INDEX − 1) + N]
where,
@TIME :  Measurement start time.
Thold :  Hold time.
Tinterval :  Sampling interval.
@INDEX :  Data index.
N :  Number of triggers between two nearest measurement points. This 
value is 0 if the sampling interval is enough longer than the 
measurement time. 
Parameter 
Name
Meanings of Parameter
@TIME Measurement start time. This is the time the measurement unit 
starts one point measurement. This is different from timing of the 
measurement trigger sent every sampling interval.
@INDEX Data index. Integer. This is the index numbered to measurement 
data stored in memory.
V1 for example, SMU1 voltage output value or measured value.
I1 for example, SMU1 current output value or measured value.