3-10 Agilent 4155C/4156C User’s Guide Vol.2, Edition 6
Sampling Measurements
Measurement Functions
To use this function, the INITIAL INTERVAL value must be set to 2 ms or more. 
The INITIAL INTERVAL is the minimum resolution of the sampling interval. For 
details about the INITIAL INTERVAL, see “Linear Sampling Measurement” on 
page 3-13, “Thinned-out Sampling Measurement” on page 3-16, or “Logarithmic 
Sampling Measurement” on page 3-19.
To set up the stop condition, specify the following parameters on the MEASURE: 
SAMPLING SETUP screen.
Parameter Description
ENABLE/ 
DISABLE  Enables or disables the stop condition.
ENABLE DELAY  Delay time. in second. This is the time from starting sampling 
measurement to enabling this function. 
Allowable range: 0 to INITIAL INTERVAL × 32767 s. 
Resolution: INITIAL INTERVAL.
NAME  Name of measurement data or user function to monitor for stop 
condition. 
Val of EVENT.
THRESHOLD  Threshold value at which to stop sampling measurement. 
Th of EVENT.
EVENT  Event for stop condition.
Val > Th  True if NAME parameter value is greater than 
THRESHOLD value.
Val < Th  True if NAME parameter value is less than 
THRESHOLD value.
|Val| > |Th|  True if absolute NAME parameter value is 
greater than absolute THRESHOLD value.
|Val| < |Th|  True if absolute NAME parameter value is less 
than absolute THRESHOLD value.
EVENT NO.  Target value of the count the event occurs (true). When the 
count of true is this value, sampling is immediately stopped. 
Allowable range: 1 to 200.