Sampling Measurements
Measurement Functions
Logarithmic Sampling Measurement 
Logarithmic sampling mode plots the measurement data on the X-axis (@TIME) set 
to the logarithmic scale by doing the following operation. See Figure 3-2.
1. Forces constant current, constant voltage, or pulsed constant bias. 
2. Waits hold time.
3. Triggers one point measurement.
4. Measurement unit executes measurement. Measurement result data is stored in 
memory.
5. Triggers one point measurement. Interval of trigger is constant (setting value of 
INITIAL INTERVAL).
6. Measurement unit executes measurement if it is ready to measure. Measurement 
result data is stored in memory.
Measurement unit waits next trigger if it is busy or in measurement.
7. Repeats steps 5 and 6 until that a sampling completion condition is satisfied. 
Logarithmic sampling mode retains only the measurement data that can plot the 
data on the logarithmic X-axis in almost the same distance.
Figure 3-2 Example Operation of Logarithmic Sampling
TIME
Hold Time
Discard
Sampling Interval
Starts
Sampling
Data of first decade
Measurement
Data of second decade
@TIME
This is the example
to plot LOG10 data.
Second decadeFirst decade
1/10 decade