Installation guide
131
Agilent 16900 Series Logic Analysis Systems
Installation Guide
A
A
Safety Notices
This apparatus has been designed and tested in accordance with IEC 
Publication 1010, Safety Requirements for Measuring Apparatus, and has 
been supplied in a safe condition. This is a Safety Class I instrument 
(provided with terminal for protective earthing). Before applying power, 
verify that the correct safety precautions are taken (see the following 
warnings). In addition, note the external markings on the instrument that 
are described under "Safety Symbols."
Warnings
• Before turning on the instrument, you must connect the protective 
earth terminal of the instrument to the protective conductor of the 
(mains) power cord. The mains plug shall only be inserted in a socket 
outlet provided with a protective earth contact. You must not negate the 
protective action by using an extension cord (power cable) without a 
protective conductor (grounding). Grounding one conductor of a 
two- conductor outlet is not sufficient protection.
• Only fuses with the required rated current, voltage, and specified type 
(normal blow, time delay, etc.) should be used. Do not use repaired 
fuses or short- circuited fuse holders. To do so could cause a shock or 
fire hazard.
• If you energize this instrument by an auto transformer (for voltage 
reduction or mains isolation), the common terminal must be connected 
to the earth terminal of the power source.
• Whenever it is likely that the ground protection is impaired, you must 
make the instrument inoperative and secure it against any unintended 
operation.