Installation guide
  81
Chapter 4: Installing Oscilloscope Measurement Modules
16533/34A Oscilloscope Module (single or multi-card modules)
10 If you are only calibrating one card set the PROTECTED/
UNPROTECTED switch back to PROTECTED. 
11 For multi-card modules, leave the switch set to UNPROTECTED and 
continue to Step 4 to perform reconfiguration.
Step 4
Reconfigure multi-card modules
A multicard module should contain either all 16533A or 16534A cards. 
1 After calibrating each card individually (following the procedures 
beginning on page 72) and before applying power to the mainframe, 
connect the module cables.
a Beginning with the top-most card, connect the ECL EXT TRIG OUT 
to the ECL EXT TRIG IN of the card immediately below. Use the 
master/expander trigger cable included with the accessory kit of 
each card.
PROTECTED
UNPROTECTED
If only calibrating one
card set protect.
Master/expander
trigger cables