Installation guide
Table Of Contents
- A Quick Tour
- Before You Begin
- Introduction to Timing Analysis: Trigger on an Edge
- Verify Pulse Widths
- Introduction to State Analysis: Trigger on an Event
- Trigger on a Sequence of Events
- Trigger on a 4 Bit Serial Pattern
- Trigger the Oscilloscope with the Timing Analyzer- Load the RESET Configuration File
- Connect the Oscilloscope Probe and Turn the Glitch On
- Get the Analog Waveform on the Display
- Set Up the Timing Analyzer
- Set Up the Timing Analyzer to Trigger on the Glitch
- Tell the Oscilloscope When to Trigger
- Set Up the Analyzer to Arm the Oscilloscope
- Run the Timing Analyzer and Oscilloscope
- Add the Analog Waveform to the Timing Waveform
- Turn the Glitch Off
- Save Your Work
- Lesson Summary
 
- Using the Pattern Generator- Load the RESET Configuration File
- Connect the Pattern Generator
- Set Up the Timing Analyzer
- Set Up the Bus Labels
- Define the Trigger Conditions: Trigger on a 1
- Set Up the Pattern Generator
- Program the Pattern Generator Output
- Start the Pattern Generator and View the Walking Ones Pattern
- Stop the Pattern Generator
- Save Your Work
- Lesson Summary
 
- Setting the Jumpers
- About the Credit Card Board

Hewlett-Packard
P.O. Box 2197
1900 Garden of the Gods Road
Colorado Springs, CO 80901-2197, U.S.A.
© Copyright Agilent Technologies  
1998 All Rights Reserved.
Reproduction, adaptation, or 
translation without prior written 
permission is prohibited, except 
as allowed under the copyright 
laws.
Restricted Rights Legend
Use, duplication, or disclosure by 
the U.S. Government is subject to 
restrictions set forth in 
subparagraph (C) (1) (ii) of the 
Rights in Technical Data and 
Computer Software Clause in 
DFARS 252.227-7013. Agilent 
Technologies, 3000 Hanover 
Street, Palo Alto, CA 94304 U.S.A.  
Rights for non-DOD U.S. 
Government Departments and 
Agencies are set forth in FAR 
52.227-19 (c) (1,2).
Document Warranty
The information contained in this 
document is subject to change 
without notice.
Agilent Technologies makes 
no warranty of any kind with 
regard to this material, 
including, but not limited to, 
the implied warranties of 
merchantability or fitness for 
a particular purpose.
Agilent Technologies shall not be 
liable for errors contained herein 
or for damages in connection with 
the furnishing, performance, or 
use of this material.
Safety
This apparatus has been designed 
and tested in accordance with 
IEC Publication 1010, Safety 
Requirements for Measuring 
Apparatus, and has been supplied 
in a safe condition. This is a 
Safety Class I instrument 
(provided with terminal for 
protective earthing). Before 
applying power, verify that the 
correct safety precautions are 
taken (see the following 
warnings). In addition, note the 
external markings on the 
instrument that are described 
under "Safety Symbols."
Warning
• Before turning on the 
instrument, you must connect the 
protective earth terminal of the 
instrument to the protective 
conductor of the (mains) power 
cord. The mains plug shall only be 
inserted in a socket outlet 
provided with a protective earth 
contact. You must not negate the 
protective action by using an 
extension cord (power cable) 
without a protective conductor 
(grounding). Grounding one 
conductor of a two-conductor 
outlet is not sufficient protection.
• Only fuses with the required 
rated current, voltage, and 
specified type (normal blow, time 
delay, etc.) should be used. Do 
not use repaired fuses or short-
circuited fuseholders.  To do so 
could cause a shock of fire hazard.
• Service instructions are for 
trained service personnel. To 
avoid dangerous electric shock, 
do not perform any service unless 
qualified to do so. Do not attempt 
internal service or adjustment 
unless another person, capable of 
rendering first aid and 
resuscitation, is present.
• If you energize this instrument 
by an auto transformer (for 
voltage reduction), make sure the 
common terminal is connected to 
the earth terminal of the power 
source.
• Whenever it is likely that the 
ground protection is impaired, 
you must make the instrument 
inoperative and secure it against 
any unintended operation.
• Do not operate the instrument 
in the presence of flammable 
gasses or fumes. Operation of any 
electrical instrument in such an 
environment constitutes a definite 
safety hazard.
• Do not install substitute parts or 
perform any unauthorized 
modification to the instrument.
• Capacitors inside the 
instrument may retain a charge 
even if the instrument is 
disconnected from its source of 
supply.
Safety Symbols
Instruction manual symbol: the 
product is marked with this 
symbol when it is necessary for 
you to refer to the instruction 
manual in order to protect against 
damage to the product.
Hazardous voltage symbol.
Earth terminal symbol: Used to 
indicate a circuit common 
connected to grounded chassis.
WARNING
The Warning sign denotes a 
hazard. It calls attention to a 
procedure, practice, or the like, 
which, if not correctly performed 
or adhered to, could result in 
personal injury. Do not proceed 
beyond a Warning sign until the 
indicated conditions are fully 
understood and met.
CAUTION
The Caution sign denotes a 
hazard. It calls attention to an 
operating procedure, practice, or 
the like, which, if not correctly 
performed or adhered to, could 
result in damage to or destruction 
of part or all of the product. Do 
not proceed beyond a Caution 
symbol until the indicated 
conditions are fully understood or 
met.
!










