Specifications
110
Chapter 1: Agilent Technologies 16700A/B-Series Logic Analysis System
Arming Second Analyzer
Arming Second Analyzer
If you are using both analyzers on the same measurement module, you 
can have one analyzer arm the other analyzer. The functionality of this 
feature is handled in two different ways, depending on the analyzer 
model.
167xx - Arming Second Analyzer
If you are using a 16715 and newer analyzer, the arming of one analyzer 
by another is handled by the trigger function "Wait for second analyzer 
to trigger".
165xx - Arming Second Analyzer
If you are using a 165xx analyzer, the arming of one analyzer by 
another is handled by the "Machine Arming Tree" found under Arming 
Control in the logic analyzer's Trigger tab - Settings tab.
The order in which the analyzers appear in the Arming Tree 
determines the order they receive the arming signal. The arm signal 
does not automatically start the next analyzer. The arm signal simply 
tells the next analyzer that it can start evaluating its own trigger 
specification and run when its trigger conditions are satisfied.
If a trigger specification in an analyzer has multiple sequence levels, 