Specifications
143
Chapter 1: Agilent Technologies 16700A/B-Series Logic Analysis System
The System Administration Tools
To terminate the telnet session
To terminate your remote session, select Exit in the remote System 
window. Then from the remote Session Manager window, select 
Disconnect. This will remove both the remote System window and 
Session Manager window from the local display.
NOTE: Hostnames that are not located in /etc/host on your local computer may not be 
recognized by the logic analysis system. If this is the case, use the Internet IP 
address instead of the hostname.
Troubleshooting the 100BaseT Lan Connection
The following troubleshooting information is for the 100BaseT LAN 
connection to the logic analysis system.
Duplex Mode 
Check the network switch duplex mode setting. The analyzers are set 
to AUTO_DETECT. The network switches should also be set to 
AUTO_DETECT. (AUTO_DETECT is also known as AUTO_Negotiate.)
If the network switch is not set to AUTO_DETECT, the analyzer will 
default to HALF_DUPLEX. This is in accordance with IEEE standards. 
Therefore, if the network switch is not set to AUTO_DETECT, it should 
be set to HALF_DUPLEX.
If the network switch is set to FULL_DUPLEX duplex mismatch will 
occur. The symptoms are a very slow data transfer in one direction and 
a relatively fast data transfer in the other direction. The network 
switch should be changed to AUTO_DETECT.
Network 
Configuration 
Verify Cabling 
Make sure UTP Category 5 is used, and that it is in good condition.
Verify that the RJ45 connector pins match the following: 
Receive Signal: Pin 1 = White, Pin 2 = Orange Transmit Signal: Pin 3 = 
White, Pin 6 = Green 
Make sure the cable length is not between 35 and 41 meters. If it is, 
then expand or reduce the length.