Specifications
152
Chapter 1: Agilent Technologies 16700A/B-Series Logic Analysis System
The System Administration Tools
selected, each individual board or module lists a set of its own 
individual tests. Some measurement modules are composed of multiple 
cards where one card is called the master and the others are called 
expanders. Self tests for a multiple card set are always tested through 
the master card. Test results for the expander cards will track results 
for the master card. In some cases, the self tests can identify a problem 
to either an expander or master card. In this case, the failure counts 
shown on an expander card may differ from those shown for a master 
card.
Test Options
Test options are set from the Options pulldown in the menu bar. They 
are provided only as a convenience to the user. They are intended for 
factory use only.
User Action Required
When running "All Tests", some of the tests may have a test result of 
"Not executed". In some cases, this message is because the test does 
not apply to the current hardware configuration. In other cases, it 
indicates the test requires some type of user action. These tests should 
be individually selected, and when user action is required, a dialog will 
appear to direct the user. The tests that require user action are listed 
below.
CPU Board 
• Floppy Drive Test: A DOS formatted floppy must be in the drive.
• External SCSI Test: A powered-up CD-ROM drive must be on the bus. 
HP16517A Timing Module 
• Skew Adjust: Not a test; used for calibration. 
HP16522A Stimulus Module 
• Output Stimulus Vectors: Not a test; provides continuous signal output. 
HP16534A Scope Module 
• ADC Test: There must be no stimulus on the scope inputs. 
HP16550 State and Timing Module 