Specifications
29
Chapter 1: Agilent Technologies 16700A/B-Series Logic Analysis System
System Overview
 149 degrees F)
 - Humidity
 Instrument, probe lead sets, and cables: 8 to 80% relative humidity at 
  40 degrees C (+104 degrees F)
 - Altitude To 3000 m (10,000 ft) 
 - Vibration 
 Operating: Random vibration 5-500 Hz, 10 minutes per axis, 
 approximately 0.2 g rms
 Nonoperating: Random vibration 5 to 500 Hz, 10 minutes per axis,
 approximately 2.41 g rms; and swept sine resonant 
 search, 5 to 500 Hz, 0.50 g (0-peak), 5-minute 
 resonant dwell at 4 resonances per axis. 
What is a Specification
A Specification is a numeric value, or range of values, that bounds the 
performance of a product parameter. The product warranty covers the 
performance of parameters described by specifications. Products 
shipped from the factory meet all specifications. Additionally, the 
products sent to Agilent Technologies Customer Service Centers for 
calibration and returned to the customer meet all specifications.
Specifications are verified by Calibration Procedures.
What is a Characteristic
Characteristics describe product performance that is useful in the 
application of the product, but that is not covered by the product 
warranty. Characteristics describe performance that is typical of the 
majority of a given product, but not subject to the same rigor 
associated with specifications.
Characteristics are verified by Function Tests.
What is a Calibration Procedure
Calibration procedures verify products or systems operate within the 
specifications. Parameters covered by specifications have a 
corresponding calibration procedure. Calibration procedures include 
both performance tests and system verification procedure. Calibration 