Specifications
68
Chapter 1: Agilent Technologies 16700A/B-Series Logic Analysis System
Overview - Multiple Instrument Configuration
and also that it knows each sample's time relationship to its own 
trigger. 
A big benefit of time correlation is the ability to display multiple data 
sets in the same display window with all data time-stamped and aligned 
relative to each other.
Timing analyzers are always referenced to the time base of the 
measurement. However, if you are using a State analyzer, you must set 
its Count field (see page 112) to Time.
Skew Adjustment
Skew is a small timing deviation between the different instruments 
configured in the same measurement. It is usually due to variances in 
internal probing delays from one instrument to another. You 
compensate for any variations using the Intermodule Skew (see 
page 101) adjustment.
See Also
For another example of a multiple instrument measurement, refer to 
Using a Timing Analyzer and an Oscilloscope. (see page 83)