Specifications
87
Chapter 1: Agilent Technologies 16700A/B-Series Logic Analysis System
Using a Timing Analyzer and an Oscilloscope
Configure the Group Run Arming Tree
How you configure the Group Run Arming Tree determines if an 
instrument tool runs as an independent tool or in a group. By default, 
all instrument tools are Independent and in Single acquisition mode.
NOTE: All multiple instrument configurations must be part of a Group Run to take 
advantage of time-correlation and intermodule arming.
• From the Intermodule window, select the oscilloscope icon and set it to be 
armed by Group Run. Select the Timing analyzer and set it to be armed by 
the oscilloscope. 
The result is an arming tree that has the scope armed when the Group Run 
field is selected. Then the timing analyzer is armed when the scope 
triggers on the glitch. The analyzer is configured to trigger on the first data 
it finds. 