Specifications
91
Chapter 1: Agilent Technologies 16700A/B-Series Logic Analysis System
Using the Correlation Dialog
Using the Correlation Dialog
Two required elements of a multiple instrument configuration are the 
Arming Signal and the Time Correlation of Data Sets. The 
Correlation dialog is used to help you correct any configuration 
problems with these two elements. If you run a measurement and the 
Correlation dialog appears, reconfigure any inputs that show errors.
• Armed From – For each Input Name, set the Armed From field to either 
the Group Run or another instrument. This allows the arming signal to 
create an ordered triggering sequence between instruments.  For more 
information, refer to "The Arming Signal", in the Overview - Multiple 
Instrument Configuration (see page 67).
• Time Tags – Each data set must be acquired in reference to a common 
time base or time reference. Timing analyzers are always referenced to the 
timebase of the measurement. However, if you are using a State analyzer, 
you must set its Count field to Time (time tags). For more information, 
refer to "Time Correlation of Data Sets", in the Overview - Multiple 
Instrument Configuration (see page 67). 