Technical data
2 MEASurement Commands
110 N1911A/1912A P-Series Power Meters Programming Guide
FETCh[1]|2|3|4 Queries
The FETCh? queries set the specified window’s measurement function. This 
can be set to either single channel, difference or ratio measurements, with 
relative mode either off or on. They then recalculate the measurement and 
place the result on the bus. The format of the result is set by 
FORM[:READ][:DATA]. Refer to Chapter 6, “FORMat Subsystem,” on page 
249 for further information.
The query returns a measurement result when it is valid. The 
measurement result is invalid under the following conditions:
•When *RST is executed.
• Whenever a measurement is initiated.
•When any SENSe parameter, such as frequency, is changed.
If data is invalid, the FETCh? query is not completed until all data 
becomes valid. The exceptions to this are, if the power meter is in the idle 
state and the data is invalid, or the power meter has been reconfigured as 
defined above and no new measurement has been initiated. In such cases, 
the  FETCh? routine generates the error –230, “Data corrupt or stale” and 
no result is returned. A common cause for this error is receiving a FETCh? 
after a *RST. If the expected value and resolution parameters are not the 
same as those that were used to collect the data, error –221, “Settings 
conflict” occurs.
NOTE
When TRIG:SOUR is INT1, INT2 or EXT and a new acquisition has been initiated (using the 
INIT command for example), FETCH? waits until the trigger takes place before executing. If triger 
conditions are not satisfied - when the trigger level differs greatly from the signal level for example - 
this can give the impression that the power meter has hung.
To unlock the power meter and adjust trigger settings, an SDC (Selected Device Clear) GPIB 
Command must be performed. This is equivalent to “EXECUTE CLEAR” in Agilent VEE.