Technical data
MEASurement Commands 2
N1911A/1912A P-Series Power Meters Programming Guide 133
Parameters
Refer to “Optional Parameters” on page 90 for additional details on the 
parameters in this command.
1
 The mnemonic DEF means DEFault. This is not equivalent to the DEFault parameter used in the 
command sub-systems. The parameters must be entered in the specified order. If parameters are 
omitted, they default from the right. The parameter DEFault is used as a place holder. Specifying 
DEF leaves the parameter value unchanged.
2
 When the measurement result is linear this parameter represents the number of significant 
digits. When the measurement result is logarithmic 1 to 4 represents of 1, 0.1, 0.01 and 0.001 
respectively.
Item Description/Default Range of Values
expected_value
(for the expected 
power level)
The expected power level parameter can be 
set to DEF or a numeric value. If a value is 
entered it should correspond to that set by 
CONFigure otherwise an error occurs.
sensor dependent
DEF
1
resolution A numeric value for the resolution. If it is 
unspecified the current resolution setting is 
used. If a value is entered it should 
correspond to the current resolution setting 
otherwise an error occurs.
1 to 4
2
1.0, 0.1, 0.01, 0.001
DEF
1
source list The channel which the command is 
implemented on. 
If unspecified the current window setup is 
used. However, on the N1912A, if the window 
shows a ratio or difference measurement, the 
upper window defaults to Channel A and the 
lower window to Channel B.
(@1)
(@2) (N1912A only)