Technical data
4 CALibration Subsystem
202 N1911A/1912A P-Series Power Meters Programming Guide
CALibration[1]|2[:ALL]
This command causes the power meter to perform a calibration sequence 
on the specified channel. The command assumes that the power sensor is 
connected to the POWER REF output. The calibration sequence consists of:
1 Zeroing the power meter (CALibration:ZERO:AUTO ONCE), and
2 Calibrating the power meter (CALibration:AUTO ONCE).
For 8480 Series power sensors and N8480 Series power sensors with 
Option CFT, the reference calibration factor used during this calibration 
can be derived from either an active sensor calibration table or the value 
entered using CALibration:RCFactor. The actual value used is the one 
which was most recently set. That is, a value entered using 
CALibration:RCFactor is overridden if a sensor calibration table is 
subsequently selected and enabled. Conversely, CALibration:RCFactor 
overrides any reference calibration factor previously set from a sensor 
calibration table. To determine the currently set reference calibration 
factor use CALibration:RCFactor?.
E- Series power sensors and N8480 Series power sensors (excluding Option 
CFT) have their sensor calibration tables stored in EEPROM which means 
that the reference calibration factor is automatically downloaded by the 
power meter.
NOTE
This command is identical to CALibration[1]|2[:ALL]?, however, unlike the query it 
does not provide a response to indicate whether the calibration has been successful or not.