Technical data
DISPlay Subsystem 5
N1911A/1912A P-Series Power Meters Programming Guide 245
DISPlay[:WINDow[1]|2]:STATe <boolean>
This command enables/disables the upper or lower window (WINDow1 and 
WINDow2 respectively) so that the display shows a single window only. The 
displayed window is presented in expanded format, showing a single 
measurement only: either the single measurement that was shown on the 
window, or the currently selected measurement, if two measurements had 
been shown. 
Syntax
Examples
Reset Condition
On reset, both windows are enabled.
DISP:WIND2:STAT OFF This command disables the lower window. 
The upper window in shown in expanded 
format, displaying its currently selected 
measurement.
DISP:WIND2:STAT ON This command enables the lower window 
so that a dual window display is once 
more provided.
:STAT
0|OFF
1|ON
Space
?
:WIND
1
2
DISP