Technical data
7MEMory Subsystem
284 N1911A/1912A P-Series Power Meters Programming Guide
MEMory:TABLe:GAIN[:MAGNitude] 
<numeric_value>{,<numeric_value>}
This command is used to enter calibration factors into the sensor 
calibration table (8480 Series sensors and N8480 Series sensors with 
Option CFT only) or offsets into the frequency dependent offset table, 
currently selected using MEMory:TABLe:SELect. Any previous calibration 
factor list, or offset list is cleared before the new calibration 
factors/offsets are stored.
A maximum of 81 parameters for sensor calibration tables and 80 
parameters for frequency dependent offset tables can be sent with this 
command. For sensor calibration tables only, the first parameter is the 
reference calibration factor, each subsequent parameter is a calibration 
factor point in the sensor calibration table.
Entries in the frequency lists correspond as shown in Table 7- 22 with 
entries in the calibration/offset factor lists.
Tab le  7- 2 2 Frequency and Calibration/Offset Factor List
For sensor calibration tables the number of frequency points must be one 
less than the number of calibration factor data points. This is verified 
when the sensor calibration table is selected using 
SENSe:CORRection:CSET1:SELect <string>.
Frequency Calibration Factor/Offset
- Reference Calibration Factor
(For Sensor Calibration Tables)
Frequency 1 Calibration Factor/Offset 1
""
Frequency 80 Calibration Factor/Offset 80