Technical data
9 PSTatistic Subsystem
334 N1911A/1912A P-Series Power Meters Programming Guide
PSTatistic[1]|2:CCDF:DATa?
This command is used to return 501 probability values in % at different 
power level within certain range, starts from 0 dB till the predefined 
maximum power level.
Syntax
Example
NOTE
The maximum power level can be set by using this command: 
PSTatistic[1]|2:CCDF:DATa:MAX <numeric_value>
By default, the maximum value is 50 dB.
The power interval between each reading (probability value) is determined by the defined 
maximum power level divided by 500.
NOTE
This command is only applicable when P-Series sensors are present and the following 
conditions are met:
• Free run acquisition mode is selected
• NORMal or DOUBle measurement speed setting is chosen
PST
2
:CCDF
:DAT
1
?
PST1:CCDF:DAT? This command returns 501 probability 
values in % at different power levels 
within certain range (from 0 dB to 
maximum power level defined).