Technical data
10 SENSe Subsystem
366 N1911A/1912A P-Series Power Meters Programming Guide
[SENSe[1]]|SENSe2:AVERage:COUNt:AUTO <boolean> 
This command enables and disables automatic averaging. ONCE has no 
affect on the power meter.
When the auto filter mode is enabled, the power meter automatically sets 
the number of readings averaged together to satisfy the averaging 
requirements for most power measurements. The number of readings 
averaged together depends on the resolution and the power level in which 
the power meter is currently operating. Figure 10- 20 is an example of the 
averaged number of readings for each range and resolution when the 
power meter is in auto measurement average mode and using a P- Series 
or E932XX power sensor. 
Setting this command to ON automatically sets the 
[SENSe[1]]|SENSe2:AVERage:STATe command to ON.
Figure 10-20Example of Averaged Readings
If  [SENSe[1]]|SENSe2:AVERage:COUNt:AUTO is set to OFF, the filter 
length is set by the [SENSe[1]]|SENSe2:AVERage:COUNt command. Using 
the  [SENSe[1]]|SENSe2:AVERage:COUNt command disables automatic 
averaging.
Auto averaging is enabled by the MEASure:POWer:AC? and 
10 dB
10 dB
Maximum Sensor Power
Minimum Sensor Power
Power Sensor
Dynamic Range
10 dB
10 dB
1234
1118
Resolution Setting
Number of Averages
11116
11232
1 1 16 256
1 8 128 128