Technical data
SENSe Subsystem 10
N1911A/1912A P-Series Power Meters Programming Guide 369
[SENSe[1]]|SENSe2:AVERage:SDETect <boolean> 
This command enables and disables step detection. In AUTO filter mode, 
the average of the last four values entered into the filter is compared to 
the average of the entire filter. If the difference between the two averages 
is greater than 12.5%, the digital filter is cleared. The filter then starts 
storing new measurement values. This feature shortens the filter time 
when the input power changes substantially. for the filter output to get to 
its final value. Note that this result appears to settle faster, although true 
settling to the final value is unaffected.
Syntax
NOTE
Step detection is automatically disabled when TRIG:DEL:AUTO is ON and the trigger 
mode is set to free run.
Under this circumstances the value of SENS:AVER:SDET is ignored. Note also that 
SENS:AVER:SDET is not set by the instrument (that is, SENS:AVER:SDET retains its 
current setting which may indicate that step detection is ON).
NOTE
With certain pulsing signals step detect may operate on the pulses, preventing the final 
average being completed and making the results unstable. Under these conditions SDET 
should be set to OFF.
SENS
1
AVER
SENS2
:
:SDET
0|OFF
1|ON
Space
?