Technical data
10 SENSe Subsystem
430 N1911A/1912A P-Series Power Meters Programming Guide
[SENSe[1]]|SENSe2:POWer:AC:RANGe:AUTO <boolean> 
This command is only valid when used with an E-Series power sensor or 
N8480 Series power sensor (excluding Option CFT). Its purpose is to 
enable and disable autoranging. When autoranging is ON, the power meter 
selects the best measuring range for the measurement. When autoranging 
is set to OFF, the power meter remains in the currently set range.
The  [SENSe[1]]|SENSe2:POWer:AC:RANGe command disables autoranging.
If  INITiate:CONTinuous is set to ON and TRIGger:SOURce is set to 
IMMediate, the range tracks the input power if 
[SENSe[1]]|SENSe2:POWer:AC:RANGe:AUTO is ON.
If the power meter is not making measurements then autoranging only 
occurs when the power meter is triggered.
Syntax
Example
NOTE
For E9320 power sensor, only UPPER and LOWER ranges are available in Normal and 
Triggered modes. 
POW:AC:RANG:AUTO 0 This command disables autoranging.
SENS
1 :AC
POW
SENS2
:
:AUTO
0|OFF
1|ON
Space
?
:RANG