Technical data
11 STATus Subsystem
470 N1911A/1912A P-Series Power Meters Programming Guide
Parameters
Query
:ENABle?
The query returns a 15 bit decimal- weighted number representing the 
contents of the Enable Register of the SCPI Register Set being queried. 
The format of the return is <NR1> in the range of 0 to 32767 (2
15
–1). 
:NTRansition <NRf>|<non-decimal numeric> 
This command sets the Negative Transition Register of the SCPI Register 
Set you require to control. The parameter value, when rounded to an 
integer and expressed in base 2 has its first 15 bits written into the 
Negative Transition Register of the SCPI Register Set concerned. The last 
bit (bit 15) is always set to 0.
Type Description Range of Values
NRf The value used to set the 
Enable Register.
0 to 2
16
–1
non-decimal numeric