Technical data
1 Power Meter Remote Operation
22 N1911A/1912A P-Series Power Meters Programming Guide
setting the filter length to 1024 and the trigger delay off.
1 ABOR2
2 CONF1 -50,DEF,(@2)
3 SENS2:AVER:COUN 1024
4 TRIG2:DEL:AUTO OFF
5 INIT2
6 FETC1? -50,DEF,(@2)
Example 4 - Specifying the Resolution Parameter
The previous examples detailed the use of the expected value and source 
list parameters. The resolution parameter is used to set the resolution of 
the specified window. This parameter does not affect the resolution of the 
data, however it does affect the auto averaging setting (refer to Figure 1- 2 
on page 35).
Since the filter length used for a channel with auto- averaging enabled is 
dependent on the window resolution setting, a conflict arises when a given 
channel is set up in both windows and the resolution settings are 
different. In this case, the higher resolution setting is used to determine 
the filter length.
The following example uses the resolution parameter to specify a 
resolution setting of 3. This setting represents 3 significant digits if the 
measurement suffix is W or %, and 0.01 dB if the suffix is dB or dBm (for 
further details on the resolution parameter refer to the commands in 
Chapter 2, “MEASurement Commands”). Also, in this example the 
expected power and source list parameters are defaulted. The expected 
power value is left unchanged at its current setting. The source list 
parameter is defaulted as described in the note “Agilent N1912A Only” on 
page 15. Note that as the source list parameter is the last specified 
parameter you do not have to specify DEF.