Technical data
12 SYSTem Subsystem
544 N1911A/1912A P-Series Power Meters Programming Guide
CDMA2000 
The cdma2000 set- up provides:
• Average power in a cdma2000 signal (bandwidth <=5 MHz).
• Peak power and peak- to- average ratio of the signal over a defined, 
statistically valid number of samples. The reading is continuously 
refreshed. This indicates how cdma2000 channel loading affects peak 
power and power distribution. 
The measurement is a continuously gated measurement on a 3 GPP 
cdma2000 signal. Its aim is to measure the peak and average power 
corresponding to a <0.01 % probability that there are no peaks above the 
returned peak reading. Time gating is set to 10 ms, corresponding to 
200,000 samples. Triggering is set to occur continuously internally to the 
meter. The internal trigger is set to AutoLevel. A reading over the 10 ms 
period is returned, then the reading is re- initiated for the next 10ms 
period. In this way the reading always relates to a position beyond 0.01 % 
on the CCDF curve and will refresh to track any signal or DUT changes.
The following table shows the power meter presets when 
<character_data> is set to CDMA2000. Commands not listed are preset 
according to their DEFault values (for further information refer to 
Table 12- 26):
Tab le  12 - 36  cdma2000: Power Meter Presets
Command Setting Comments
Frequency
[SENS[1]]|SENS2:FREQ[:CW|:FIX] +1900.000 MHz Frequency setting
Sensor measurement mode
[SENS[1]]|SENS2:DET:FUNC NORM Measurement mode
Sensor video bandwidth setup
[SENS[1]]|SENS2:BAND|BWID:VID E9321A/25A: DEF
E9322A/26A: DEF
E9323A/27A: OFF
N1921/2A: OFF
Sensor video bandwidth
Gate Setup
[SENS[1]]|SENS2:SWE[1]|2|3|4
:OFF:TIME
Gate 1: 0 s
Gates 2 - 4: 0
Delay between trigger point and time 
gated period. 