Technical data
TRACe Subsystem 13
N1911A/1912A P-Series Power Meters Programming Guide 613
Parameters
Example
Error Messages
If  TRAC:STAT is off, the error –221, “Settings Conflict” occurs. 
Item Description/Default Range of 
Values
character_data • HRESolution: high resolution. The complete capture 
buffer at the internal sample rate. The number of points in 
this trace is not fixed, as it is affected by the 
SENS:TRACe:TIMe setting. 
• MRESolution: medium resolution. A subset of the 
capture buffer - the buffer contents are decimated
1 
to 1000 
data points. 
• LRESolution: low resolution. A subset of the capture 
buffer - the buffer contents are decimated
1
 to provide 230 
data points. This is the same number of data points as the 
power meter uses to display the trace on the front panel. 
Hence, the LRES command can be used to replicate the 
power meter’s display.
HRES
MRES
LRES
TRAC:DATA? HRES This command returns the trace data for 
Channel A at high resolution. 