Technical data
TRACe Subsystem 13
N1911A/1912A P-Series Power Meters Programming Guide 621
present, otherwise Error –241, "Hardware Missing" is generated. 
• If free run acquisition is selected or sensor average mode selected, 
Error –221 "Settings Conflict" occurs.
NOTE
If you attempt to measure a pulse out of the range of the capture, for example, measure the 
5th pulse and there are only 4 pulses displayed, the power meter returns #0##9.91E37 as 
the result.