TRACe[1]|2:MEASurement:PULSe[1]|...|10:DURation?
This command returns the difference between a pulse and next transition 
occurrence instants. As power pulses are by definition positive pulses, the 
pulse duration is the time difference between positive and negative 
transitions of one pulse.
Algorithm
If the first transition in the trace is positive,
then
PULSe:DURation =  time the first negative transition occurs -time the first 
positive transition occurs
else
PULSe:DURation = time the second negative transition occurs -time the 
first positive transition occurs.
Syntax
Example
TRAC
:MEAS
1
2
:PUL
:DUR
?
1
S
2|...|10
TRAC2:MEAS:PULS3:DUR? This command returns the duration of 