Technical data
13 TRACe Subsystem
634 N1911A/1912A P-Series Power Meters Programming Guide
TRACe[1]|2:MEASurement:TRANsition[1]|...|10:POSitive:OCCu
rrence?
This command returns the position, relative to the trigger instant, of the 
n
th
 occurrence of a positive transition found on a trace.
Syntax
Reset Condition
On reset, this parameter is not affected.
Example
Error Messages
• If a P- Series sensor is not connected, error –241, “Hardware missing” 
occurs.
• If a P- Series sensor is connected and Free Run trigger acquisition is 
selected, error –221, “Settings conflict” occurs. 
TRAC
:MEAS
1
2
:TRAN
:POS
:OCC
?
1
2|...|10
TRAC2:MEAS:TRAN:POS:OCC? This command returns the position, 
relative to the trigger instant, of the 1st 
occurrence of a positive transition 
found on trace 2.