Technical data
TRIGger Subsystem 14
N1911A/1912A P-Series Power Meters Programming Guide 677
TRIGger[:SEQuence[1]|2]:COUNt <numeric_value> 
This command controls the path of the trigger subsystem in the upward 
traverse of the wait for trigger state. COUNt loops through the event 
detection/measurement cycle are performed. That is, COUNt measurements 
are performed in response to COUNt trigger events. 
COUNt can be set to a value >1 only when:
• [SENSe[1]]|SENSe2:MRATe <character_data> is set to FAST
• TRIGger[1]|2:SOURce set to BUS,  IMMediate or HOLD. 
When  COUNt is set to a value >1,
• CALibration[1]|2:ZERO:AUTO will switch to OFF automatically. It will 
restored to its default setting when the COUNt is set to 1.
• Setting a channel from FAST mode to NORMal mode or DOUBle mode 
will also restore both the CALibration[1]|2:ZERO:AUTO and  COUNt to 
its default setting automatically.
Syntax
:COUN Space
?
numeric_value
DEF
TRIG
:SEQ
1
2