Technical data
14 TRIGger Subsystem
684 N1911A/1912A P-Series Power Meters Programming Guide
Parameters
Example
Item Description/Default Range of Values
source Available trigger sources: 
• BUS: the trigger source is the group execute trigger 
<GET> bus command, a *TRG common command or 
the TRIGGER:IMMediate SCPI command. 
• EXTernal: the trigger source is the trigger input in the 
back panel.
• HOLD: triggering is suspended. The only way to trigger 
the power meter is to use TRIGger:IMMediate.
• IMMediate: the trigger system is always true. If 
INITiate:CONTinuous is ON the power meter is 
continually triggering free (free run mode). If an 
INITiate:IMMediate command is sent a 
measurement is triggered then the power meter 
returns to the idle state.
• INTernal: either INT1 (Channel A) or INT2 
(Channel B).
BUS
EXTernal
HOLD
IMMediate
INTernal[[1]|2]
NOTE
The trigger source is set to IMMediate on instrument power-up and when entering local 
mode.
The MEASure and CONFigure commands automatically set the trigger source to 
IMMediate.
The READ? or MEASure commands should not be used if the trigger source is set to BUS 
or HOLD. 
TRIG:SOUR IMM This command configures Channel A for 
immediate triggering.